CAESR Facility

originalcaesr2

CAESR EPR User Spectrometers

(1) EMXmicro: X-band CW

(2) E580: X-/Q-band CW/pulsed

(3) E680: X-/W-band CW/Pulsed

Development Spectrometers

(4) E380: X-band pulsed/CW, home-built AWG in progress

(5) HF EPR Instruments, Inc., 130 GHz: D-band pulsed/CW

X-band CW EPR (Bruker EMXmicro)

emxasinstalled

• microwave frequency 9.1-9.9 GHz (0.32 cm-1, 39.6 µeV, λ = 32.6 mm)

• temperature 3-300 K

• field range 0-0.6 T

• resonators 1-4

X-/Q-band (AWG) pulsed/cw (Bruker E580)

e580 here small

Arbitrary Waveform Generator pulse formation, 1.6 GHz, 0.625 ns time step

At X-band, 1kW 

• microwave frequency 9.1-9.9 GHz (0.32 cm-1, 39.6 µeV, λ = 32.6 mm)

• temperature 2–300 K 

• magnetic field 0-1.55 T

• resonator 1-5 (rectangular cavity, dielectric ring, dielectric ring ENDOR, split-ring) 


At Q-band, 140W 

• microwave frequency 33-35 GHz (1.13 cm-1, 140 µeV, λ = 8.8 mm)

• temperature 3.8-300 K 

• magnetic field 0-1.55 T 

• resonators

X-/W-band (AWG) pulsed/cw (Bruker E680)

originalcaesr

At X-band, 1.5 kW 

• microwave frequency 9.1-9.9GHz (0.32 cm-1, 39.6 µeV, λ = 32.6 mm)

• temperature 2–300 K 

• magnetic field 0-1.55 T 

• resonator 1-5 (rectangular cavity, dielectric ring, dielectric ring ENDOR, split-ring) 

At W-band, 200 mW 

• microwave frequency ~ 94 GHz (3.14 cm-1, 389 µeV, λ = 3.2 mm)

• temperature 3.8-300 K 

• magnetic field 0-6 T 

• resonator 6

ITC503 temperature status cam

Resonators

resonator en5107d2 small

Some useful information about the resonators available and the tubes and filling heights is given below. 

X-Band CW Resonators 

1. (2x) X-band Super High Sensitivity Probehead (~10 GHz, 4 – 375 K) 

• Max. Sample Access 10 mm 

2. X-band EMX High Sensitivity Probehead (~10 GHz, 4 – 375 K) 

• Max. Sample Access 10 mm 

3. X-band Dual Mode Probehead (~10 GHz, 4 – 375 K) 

• Max. Sample Access 10 mm 

4. X-band Standard TE102 Probehead (~10 GHz, 4 – 375 K) 

• Max. Sample Access 10 mm 

X-band CW/Pulse Resonators 

5. (2x) X-band Pulse-ENDOR Dielectric Resonator (EN 4118X-MD4-W1, ~9.7 GHz, 2-300K) 

• Max. Sample Access 4 mm (use <3.8 mm tubes ONLY) 

• Sample centre 55 mm (tube longer that this) 

• Resonator length 13 mm (filling height) 

• CW or pulse mode 

• Optically transparent 200 nm – 5 μm 

6. X-band Split-Ring Resonator (ER 4118X-MS2-W1, ~9.6 GHz, 2–323 K) 

• Max. sample access 2 mm (use 1.6 mm tubes) 

• Sample centre 39 mm (tube longer that this) 

• Resonator length 6 mm (filling height) 

• CW or pulse mode (tπ/2 = 16 ns, Pmw=18 dB) 

7. (2x) X-band Split-Ring Resonator (ER 4118X-MS3(1x:)-W1, ~9.6 GHz, 2–323 K) 

• Max. sample access 3 mm (use < 2.8 mm tubes) 

• Sample centre 39 mm (tube longer that this) 

• Resonator length 6 mm (filling height) 

• CW or pulse mode (t_π/2 = 16 ns, Pmw = 18 dB) 

8. (2x) X-band Dielectric Resonator (ER 4118X-MD5-w1, ~9.7 GHz, 2–323 K) 

• Max. sample access 5 mm (use < 4.8 mm tubes) 

• Sample centre 38 mm (tube longer that this) 

• Resonator length 13 mm (filling height) 

• CW or pulse mode 

• Optically transparent 200 nm – 5 μm 

W-band Resonator 

9. W-band Pulsed ENDOR Probe (EN600-1021H TeraFlex, ~94 GHz, 4-300 K) 

• Max. sample access 0.9 mm 

• Resonator filling length c.a. 3 mm 

• CW or pulse mode, 2x ENDOR resonators

10. W-band CW-EPR Probe (~94 GHz, 4-300 K) 

Sample Preparation Laboratory

caesr wetlab small

1. Schlenck line 

• operating at 10^-4 mbar 

• EPR tubes connectors for tubes with OD of 0.9 mm, 3.0 mm, 3.8 mm or 4 mm (allows you to take your EPR tube from i.e. a glovebox, to the vacuum line without contact to air) 

2. Fume cupboard 

3. Liquid nitrogen storage dewar for EPR samples 

4. Storage Fridge/freezer 

5. Work benches 

6. Analytical balance 

7. Cary60 UV-Vis spectrophotometer

8. MBraun LabStar N2 Glovebox with -30 deg. freezer, 4x BNC port

Lasers 

lasers small

There are 2-3 OPO-LASERs at 20 Hz and 50 Hz, covering 213-2200 nm and up to 60 mJ.  There are Surelite and Minilite Nd:YAG LASERs within CÆSR. For additional information, contact Dr. Kevin Henbest